CVT ISO1716 Tester is the most common device for measuring the heat of combustion or calorific value of material by putting the specimen into the oxygen cell.
CVT Building Material Calorific Value Testing ISO1716 Tester
Introduction:
Gross heat of combustion of materials is an important parameters to evaluate building materials heat release and fire base data;
CVT Calorific Value Testing ISO1716 Tester is the most common device for measuring the heat of combustion or calorific value of material by putting the specimen into the oxygen cell.
Our CVT Calorific Value Testing ISO1716 Tester is designed and manufactured as per GB/T14402-2007 and ISO1716 standards, which is suitable for calorific value of building materials testing.
CVT Calorific Value Testing ISO1716 Tester is stable, high accurate and easy for operation.
Standards:
ISO: ISO 1716
Features:
Adopt American advanced embedded technology, 7-inch large-screen color touch screen operation, and integral installation method to ensure the firmness of the touch screen.
CRT Material makes the touch screen more durable.
Direct operation, without external computer, which is convenient for use.
Automatic water injection, no need to adjust the water temperature, just install the aerobic bomb into the barrel, and the instrument can automatically do all tests.
With fault diagnosis function and built-in reminder will remind whether the ignition is successful.
Using special foam insulation technology, it is not affected by external temperature during the experiment.
Heat capacity repeatability≤0.2%, precision <0.1%, temperature resolution 0.0001K.
Reasonable structure, reliable performance, low failure rate, high precision, easy installation and maintenance.
Parameters:
Item Name | CVT Building Material Calorific Value Testing ISO1716 Tester |
Heat capacity | 10000J/K |
Temperature range | (4.5-50)℃ |
Bomb volume | 300±50ml |
Inner cylinder volume | 2000ml |
Stirring speed | 375r/min |
Stirring power | 3W |
Stirring method | Motor directly inserted into inner cylinder for mixing |
Oxygen bomb sealing | No leakage under 3.0-3.5MPa |
Oxygen bomb withstand pressure | 21MPa |
Test time | <15min |
Timer accuracy | 0.5s |
Repeatability | ≤60(J/K) |
Power supply | AC220V±10%, 50Hz |